Bowie & Dick Test Sheets

Air leaks in the sterilization chamber threatens the sterilization safety. Daily use of Bowie & Dick Test Sheets help to determine sterilizer malfunctions at an early stage. Bowie & Dick Test Sheets can be placed inside a pack made of dry towels according EN 285 standard. In case of sterilizer malfunction, the test sheet will show a non-uniform color develop-ment and possible failures can be interpreted easily by enclosed color change card.
  1. Water based, non-toxic chemical indicator comply to ISO 11140-1
  2. Test sheet can be kept for recording purpose
  3. Manufactured in absence of lead or other heavy metals

 
RELATED STANDARS

Bowie Dick Test Packs fulfill the performance requirements of ISO 11140-4.Indicator used is classified as class 2 according to ISO 11140-1 standard. CE marking is not applicable for indicator products, because this product is not considered a medical device under the clause of the EC-directive 93/42/EEC

INTENDED USE

Bowie Dick Test Packs are designed for daily monitoring of air removal and steam penetration in pre-vacuum steam sterilizers.The Bowie Dick test Packs provide immediate visual evidence of inadequate air removal or air leak, vacuum pump failures, presence of non-condensable gases, high temperature, steam penetration, wet steam problems.

OPERATION CONDITIONS

Bowie Dick Test Packs are designed to operate in pre-vacuum steam sterilizers at 121 °C 10 minutes, 134 °C 3.5 minu- tes operation conditions. The color change should be che-cked with the sheet in the back of packaging to show if the parameters have been reached.
Name:  Bowie & Dick Test Sheets.